EasyManua.ls Logo

Amoul E6 - Environmental Specifications; Power Specifications

Amoul E6
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
H-1.601.00149-2.0
Page
49
Host
Input
DC 10-28V
10.5 Patient
Description
Types
Suitable patients
Adult patients with:
sternal height from 165 to 305 mm
Maximum chest width 450 mm
Usage of E6 is not subject to patient’s
weight limit.
10.6 Pressing parameters
Description
Parameters
Pressing depth
30-53mm,error ±3mm.
Default setting: 50mm,continuous adjustable
Pressing frequency
110 times per minute,error ±3 times
Press/Release ratio
50%,erro ±5%
Pressing mode (operator
choose)
15:2 (press 15 times,and 6 secs for
ventilation)
30:2 (press 30 times,and 6 secs for
ventilation)
Continuous pressing
10.7 Battery physical specification
Description
Parameters
Size length x width x
height)
115mmⅹ80mmⅹ52mm
Weight
0.5kg
Type
lithium polymer battery
Capacity (each battery)
2900 mAh 75Wh
Voltage
25.9V
Working time (standard
patient)
≥90min(for 2 batteries)
Charging time
less than 4 hours (22°C)
Battery change
It is recommended to replace the battery
every 3 years or 500 times (full charge and
discharge)