8 Maintainance
8.1 Regular check
After each week and after each use of E6 cardiopulmonary resuscitator, do
the following:
1. Make sure the equipment is clean.
2. Make sure to install a new suction cup.
3. Make sure the patient fixing belt is attached.
4. Make sure that the two leg straps of the stabilizing belt are wrapped
around the legs.
5. Press up and release the lock button to ensure that the lock can be opened
and reset.
6. Press ON/OFF to have E6 perform a self-test to ensure that it passes the
self-test.
7. After turning the power on, observe the battery level to make sure the
battery is fully charged.
8. Press On/Off to turn the E6 power off again.
9. Make sure the external power cord (optional accessory) is intact.
Warning:
• [Electric Shock] If the external power cord (optional accessory) is
damaged, remove it and replace it immediately to avoid electric shock or
fire hazard.
8.2 Uninstall and install patient fixing belt
Uninstall:
1. Open the patient fixing belt and pull it out of the corresponding hole in
the E6 leg.
Clean as described in Section 6.1.
Install:
1. Pass the patient fixing belt through the corresponding hole on the E6 leg.