H-1.601.00149-2.0
Page
40
Start (15:2, 30:2)
When you press this button, the E6 performs 30 (or 15) chest compressions
and then pauses for 6 seconds. During the pause, the operator can perform
two ventilations. After the pause, machine repeats the pressing. At each
pause in ventilation, the device will give the voice “ventilate”twice to alert
the operator.
5.10.3 Use in radioactive laboratory
E6 can be used in a radiology laboratory. In addition to a pressing
transmission mechanism, it is primarily radiation-shielded, allowing X-ray
projection.
5.11 Remove E6 from patient
1. Press the On/Off key for 3 seconds to turn off the power.
2. If the E6 stabilizing belt is connected to the E6, remove the buffer belt
that is part of the stabilizing part from the E6 leg
3. Grasp the legs as shown below and press the lock button to remove the
upper part from the back panel.
Fig. 43 Remove E6
4. Remove the back plate if patient condition allows