H-1.601.00149-2.0
Page
51
11 EMC
Note:
E6 cardiopulmonary resuscitator machine meets the requirements of YY0505
standard electromagnetic compatibility.
Users should install and use the electromagnetic compatibility information
provided by the random file.
Portable and mobile RF communication equipment may affect the performance
of the E6 cardiopulmonary resuscitator machine, avoiding strong
electromagnetic interference when used, such as near mobile phones, microwave
ovens, etc.
The guide and the manufacturer’s statement are in the attachment.
Warning:
E6 cardiopulmonary resuscitator machine should not be used close to or stacked
with other equipment. If it must be, it should be observed to be able to operate
normally in its configuration.
In addition to the cables sold by the manufacturer of the E6 cardiopulmonary
resuscitator machine as spare parts for internal components, the use of
extra-standard accessories and cables may result in E6 increased working or
reduced immunity.
E6 can be used in the following specific EMR environment, in which user shall ensure to
operate this equipment.
Radio frequency radiation
(CISPR 11) (GB4824)
E6 generates radio frequency energy only
when operating its internal functions.
Therefore, this ventilator emits very small
amount of radio frequency radiation and
it is unlikely to cause any EMI to
electronic equipment nearby.
Radio frequency radiation
(CISPR 11) (GB4824)
E6 is applicable in all facilities, including
domestic and public LV power supply
network directly connected to house.
Harmonic wave radiation
(GB 17625.1)
Voltage fluctuation and flicker
emission
(GB 17625.2)