EasyManua.ls Logo

Amoul E6 - System Alarm; System Prompts

Amoul E6
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
H-1.601.00149-2.0
Page
43
High
priority
Pressing
motor
abnormal
Red LED light
lit
intermittently
Voice:
system error
Sound:
Dididi--didi
Pressing
motor
abnorma
l
Re-start
machine
High
priority
EEPROM
abnormal
Red LED light
lit
intermittently
Voice:
system error
Sound:
Dididi--didi
CB’s
EEPROM
error
Re-start
machine
High
priority
displacement
sensor A
abonormal
Red LED light
lit
intermittently
Voice:
system error
Sound:
Dididi--didi
displace
ment
sensor
not
connect
ed well
Change
displacement
sensor A
High
priority
displacement
sensor B
abonormal
Red LED light
lit
intermittently
Voice:
system error
Sound:
Dididi--didi
displace
ment
sensor
not
connect
ed well
Change
displacement
sensor B
High
priority
Battery is
running out
Red LED light
lit
intermittently
Voice:
Battery is running
out
Sound:
Dididi--didi
Battery
level is
less than
5%
Connect
adapter or
change
battery
Middle
priority
Battery level
is insufficient
Yellow LED
light lit
intermittently
Voice:
Check battery
status
Sound:
di-di-di
Battery
level is
less than
20%
Connect
adapter or
change
battery
Note: Acoustic signal characteristics (one '-' means 100ms)
[1]: (di-di-di-----di-di-----di-di-di-----di-di); pulse duration 200ms, alarm
period 5s
[2]: (di-di-di); the pulse duration is 200ms and the alarm period is 5s.
7.4 System prompts
Prompts
Purpose
Trigger actor

Related product manuals