7 Faults and trouble shooting
If the problem can’t be solved immediately, please contact the
manufacturer Ambulanc (Shenzhen) Tech. Co., Ltd., or the distributor
authorized by Ambulanc (Shenzhen) Tech. Co., Ltd. Please do not continue to
use the machine to avoid unnecessary damage.
7.1 Technical faults
Fault
Cause
Remedy
E6 can't start
E6 is faulty
Give to manufacturer or
authorized distributor
Battery is finished
Charge the battery
E6 can’t close
Wrong operation
Long press “On/Off” for at least
3 seconds
Power indicator unstable
Power cable not
connected well
Re-connect power cable
7.2 Physiological Alarm
Priority
LED Alarm
LED indication
Sound Alarm
Cause
Method
High
priority
Insufficient
compression
depth
Red LED light
lit
intermittently
Voice: insufficient
compression
depth, please
re-adjust the
initial position
Sound:
Dididi--didi
Patient
too
small
stop using
the E6 and
use manual
compression
7.3 System Alarm
Priority
LED Alarm
LED indication
Sound Alarm
Cause
Method
High
priority
Communicati
on error
Red LED light
lit
intermittently
Voice:
system error
Sound:
Dididi--didi
main
board
commun
-cation
wrong
check MB
and CB
connection
or looseness