H-1.601.00149-2.0
Page
52
EMI Statement - Requirements for All Equipment and Systems
E6 can be used in the following specific EMR environments, and the user shall ensure to
operate this equipment in the following EMR environments.
Contact discharge:
±8kV
Air discharge:
±15kV
Contact
discharge: ±8kV
Air discharge:
±15kV
The ground shall be of
wood, concrete or
ceramics. In case of
composite paving material,
the relative humidity shall
be at least 30%.
To power cable:
±2kV
To long I/O
cable: ±1kV
To power cable:
±2kV
To long I/O
cable: ±1kV
Power supply grade shall
be minimally the grade for
typical commercial or
medical environment.
Power frequency
magnetic field
(50/60Hz)
(GB/T 17626.8)
Power frequency magnetic
field shall be of the
horizontal characteristics as
in typical commercial or
medical environment.
Voltage sag,
short
interruption and
variation
(GB/T 17626.11)
<5%
U
T
(>95%
fall,
U
T
), 0.5 cycle;
<5%
U
T
(>95%
fall,
U
T
), 0.5 cycle;
Power supply grade shall
be minimally the grade for
typical commercial or
medical environment. It is
recommended to use UPS
to ensure continuous
operation of this product
even in case of AC power
outage.
40%
U
T
(60% fall,
U
T
), 5 cycles;
40%
U
T
(60% fall,
U
T
), 5 cycles;
70%
U
T
(30% fall,
U
T
), 25 cycles;
<5%
U
T
(>95%
fall,
U
T
), 5s;
70%
U
T
(30% fall,
U
T
), 25 cycles;
<5%
U
T
(>95%
fall,
U
T
), 5s;
Note:
U
T
refers to the AC network voltage before the test voltage is applied.
Guide and Manufacturer Statement – EMI
E6 is intended for the following EMI environments, and E6 purchaser or user shall ensure
to operate E6 in these EMI environments: