corresponding volume of transmitter frequency, where p is the maximum rated output
power in (Watt) of transmitter provided by its manufacturer.
Note 1:
For frequency of 80MHz and 800MHz, a formula in respect of high frequency should be
used.
Note 2:
ISM bands between 150kHz and 80MHz are 6.765MHz~6.795MHz,
13.553MHz~13.567MHz, 26.957MHz~27.283MHz and 40.66MHz~40.70MHz.
Note 3:
Additional factor 10/3 is used for calculation of recommended distance to the
transmitter within frequency ranges of 150kHz ~ 80MHz and 80MHz~2.5GHz, so as to
reduce the possibility of interference resulted from mobile/portable communication
devices which are accidentally taken into patient's location.
Note 4:
As EM transmission is affected by absorption and reflection of buildings, objects and
human bodies, these guidelines may not be applicable to all circumstances.
Basic EMC Properties of E6 Cardiopulmonary resuscitator
The E6 cardiopulmonary resuscitator can work normally according to the parameter
settings. See Chapter 5 of the manual for details. Alarms can be issued based on
real-time monitoring of the E6 status, and to ensure the accuracy of the parameters
below of the E6 in the electromagnetic compatibility environment:
Pressing depth
30-53mm,error ±3mm.
Default setting: 50mm,continuous adjustable
Pressing frequency
110 times per minute,error ±3 times
Press/Release ratio
50%,erro ±5%
Pressing mode (operator choose)
15:2 (press 15 times,and 6 secs for ventilation)
30:2 (press 15 times,and 6 secs for ventilation)
Continuous pressing
Parameters of EMC Cable Material
Adapter input power cable
3.0±0.01m
Adapter output power cable
1.2±0.05m