Page 12
Introduction
1.3.3 - Regular checks
The state of the product should be checked regularly, especially the following:
• Theintegrityoftheenclosure
• Anywaterhosesused
• TheAC/DCExternalPowerSupply
• Themainscable
NOTE: Do not use equipment that is damaged.
1.3.4 - Annual electrical safety checks
It is advisable to check the integrity of the insulation and protective earth of the product on an annual basis, e.g. U.K.
PAT testing.
NOTE: Do not use equipment that is damaged
1.3.5 - Replacement parts
A PS-20 (see page 20), which is the only external power supply recommended for use with the iXon camera, has been
supplied to you. If this unit fails or is damaged, please contact Andor for a replacement. Depending on the Terms
and Conditions of your Warranty, you may be charged for this replacement.
1.3.6 - Fuse replacement
The camera itself does not have a fuse. However, if a U.K. (BS 1363) mains lead has been supplied, it contains a fuse,
whose characteristics are as follow:
• Rated Current: 5 A
• Rated Voltage: 240 Vac.
• Type: BS 1362
• Size: 0.25 × 1 inch
1.3.7 - Overheating
Careshouldbetakentoensurethatthecameradoesnotoverheat,asthiscancausesystemfailure.Overheatingmay
occur if either of the following situations arises:
• Theairventsonthesidesofthedetectorheadareaccidentallyblockedorthereisinsufcientornowaterow
• Theambientairtemperatureishigherthan30ºC
To protect the camera from overheating, a thermal switch has been attached to the heat sink. If the temperature of the
heat sink rises above predened limit, the power supply to the cooler will cut off and a buzzer will sound. The cut-out will
automatically reset once the head has cooled. It is not recommended that you operate in conditions that would cause
repeated cut-outs as the thermal switch has a limited number of operations.
NOTE: When using water cooling, always use water that is above the dew point of the ambient environment
otherwise condensation may occur (please see page 14).