iXon 3
Features & Functionality
The second type of keep clean is executed between individual scans in a kinetic series, and is relevant to Non-Frame
Transfer Mode combined with either Internal or Software Trigger. It is called the Internal Keep Clean Cycle. When the
user congures a kinetics series acquisition as well as dening the exposure time and the readout mode, they also dene
the number of scans to capture and the time between the scans. During the time between individual scans the sensor
mustbekeptfreeofchargetoensurethedatacapturedisatruereectionofthelightthatfellonitduringtheexposure
period. The keep clean cycle run during this time is very similar to that described in the Idle Keep Clean Cycle on the
previous page in that the cycle is one vertical followed by a series of horizontals. In this mode however the number of
times the cycle is repeated is determined by the cycle time set by the user. The keep clean is completed with a sufcient
number of vertical shifts to ensure both the image and storage areas are charge free.
Figure 26: Internal Keep Clean Cycle