Re marks: Se lects the cal fac tor ta ble or com bi na tion of ta bles to be used and auto -
mati cally up dates the sensor.
Ex am ple: CFU SEL A,13
Se lects the fac tory ta ble plus user ta ble 3 in sen sor A.
CFUTBL Num ber of cal fac tor ta bles in the sen sor
SEN SOR
Syn tax: CFUTBL <s>
s: A or B
Re marks: Re turns the number of cal fac tor ta bles avail able in the se lected sen sor.
CFUUSE Num ber of cal fac tor ta ble be ing used
SEN SOR
Syn tax: CFUUSE <s>
s: A or B
Re marks: Re turns a number in di cat ing the cal fac tor ta ble, or com bi na tion of ta -
bles, be ing used by the se lected sen sor. Pos si ble returned val ues are:
0 = fac tory de fault ta ble
1 to 10 = user ta ble be ing used
11 to 20 = fac tory ta ble + user ta ble be ing used
CFUVLD Valid cal fac tor ta ble check
SEN SOR
Syn tax: CFUVLD <s>,<ta ble number>
s: A or B
ta ble
number: 1 to number of ta bles sup ported by the sen sor type
Re marks: Re turns a TRUE if the ta ble number passed is a valid ini tial ized ta ble
for the se lected sensor. Re turns a FALSE if it is not.
CFVAL Current cal fac tor value
SEN SOR
Syn tax: CFVAL <s>
ML2400A OM 6-27
GPIB OPERATION ML24XXA NATIVE COMMANDS