B dB (rel)
Re marks: Set to dB units in rela tive mode us ing the pres ent rela tive ref er ence
value.
C dB (ref)
Re marks: Set to dB units in rela tive mode us ing the pres ent rela tive ref er ence
value, and en able the ap pli ca tion of the cal fac tor.
D dBm
Re marks: Set units to dBm.
EMUL Select emu la tion mode
Syn tax: EMUL <mode>
mode: ML24XX (Anritsu ML2400A Se ries na tive mode)
HP436A (Hewlett-Packard)
HP437B (Hewlett-Packard)
HP438A (Hewlett-Packard)
ML4803 (Anritsu ML4803A Se ries)
Re marks: Sets the GPIB emu la tion to emu late other types of power me ters. This
com mand is avail able in any emu la tion mode, and re sets the whole
GPIB in ter face when the emu la tion mode is changed.
When se lect ing GPIB emu la tion modes, the in stru ment con fig ures it -
self to the pre set con di tions of the in stru ment to be emu lated. For ex -
am ple, when se lect ing HP 438A emu la tion, the front panel menus pass
through the pre sets for the HP 437B (which pre sets sen sor A to dBm)
then se lects HP 438A emu la tion (which pre sets sen sor A to Watts).
NOTE
This com mand re quires a white space be tween the com -
mand header (EMUL) and the pa rame ter <mode>. This
is an ML2400A-specific com mand that does not con form
to the HP Emu la tion com mand for mat de fined at the be -
gin ning of this sec tion.
H Hold mode
Re marks: Sets both chan nels to trig ger hold mode. The power meter does not trig -
ger un til it re ceives an I or T com mand. If it re ceives an R or V com -
ML2400A OM 6-99
GPIB OPERATION HP 436A EMULATION COMMANDS