HP438A (Hewlett-Packard)
ML4803 (Anritsu ML4803A Se ries)
Re marks: Set the GPIB emu la tion to emu late other types of power me ters. This
com mand is avail able in any emu la tion mode, and re sets the whole
GPIB in ter face when the emu la tion mode is changed.
When se lect ing GPIB emu la tion modes, the in stru ment con fig ures it self
to the pre set con di tions of the in stru ment to be emu lated. For ex am ple,
when se lect ing HP 438A emu la tion, the front panel menus pass through
the pre sets for the HP 437B (which pre sets sen sor A to dBm) then se -
lects HP 438A emu la tion (which pre sets sen sor A to Watts).
EN En ter com mand
Syn tax: EN
ET En ter data for a cal fac tor ta ble
Syn tax: ET<ta ble_number><freq_value><cal fac tor><ter mi nator>
ta -
ble numbe
r:
0 to 9 (F for fac tory ta ble al lowed when us ing to read a ta ble)
freq value: cal fac tor en try fre quency value
cal fac tor: cal fac tor value in percentage
ter mi na tor: EN to ter mi nate and en try
EX to ter mi nate ta ble entries
Re marks: Since the ML2400A stores the cal fac tor ta ble in for ma tion in the sen sor,
this data must be saved to the sen sor by us ing the ‘EX’ com mand or an
ad di tional com mand ‘SV’, or the data could be lost. The sav ing of the
cal fac tor ta ble data to the sen sor can be done at the end of all up dates
to a par ticu lar ta ble.
EX Exit cal fac tor ta ble mode
Syn tax: EX
Re marks: Used on the ML2400A to force a save of the cal factor table to the
sensor if the data has changed.
ML2400A OM 6-109
GPIB OPERATION HP 437B EMULATION COMMANDS