Ex am ple: To set the fre quency of the in put sig nal to 300 MHz:
FR300MZ
GT Set group trig ger
Syn tax: GT<mode>
mode: 0
1
2
Re marks: 0 = Ig nore Group Exe cute Trig ger (GET) com mand
1 = Trig ger im me di ate re sponse to ‘GET’ com mand
2 = Trig ger with de layed re sponse to ‘GET’ com mand
The GTn com mand con fig ures what the de vice does when it re ceives the
‘GET’ com mand.
For ex am ple: GT1 sets the’ GET’ (Group Exe cute Trig ger) to per form a
TR1 type trig ger.
Re lated
Com mands: TR
ID Re turn iden ti fi ca tion string
Syn tax: ID
Re marks: Re turned for mat:
<com pany name>,<model>,<firm ware ver sion>
IDN? HP 437B iden tity re quest
Syn tax: IDN?
Re marks: Re turned for mat:
<com pany name>,<model>,<firm ware ver sion>
KB Cali bra tion fac tor
Syn tax: KB<val><ter mi na tor>
val: 1.0 to 150.0%
ML2400A OM 6-111
GPIB OPERATION HP 437B EMULATION COMMANDS