val: 1 to 10
Re marks: The ML2400A Se ries can store up to 10 in stru ment con figu ra tions for
con ven ient re call. The con figu ra tion pa rame ters stored are the same pa -
rame ters the ML2400A Se ries stores in its own *SAV and *RCL na tive
com mands. There fore, RC is equiva lent to *RCL, and ST is equiva lent to
*SAV.
Se lect ing Reg is ter 0 al ways re stores the pre vi ous power me ter con figu -
ra tion, pro vid ing an ex pe di ent way to re cover from an en try er ror.
RE Set deci mal point reso lu tion
Syn tax: RE<number>EN
val: 1, 2, or 3
Re marks: Set the number of deci mal places dis played.
Ex am ple: To set the dis play reso lu tion to 2 deci mal places:
RE2EN
RF Set the ref er ence cal fac tor value for a ta ble
Syn tax: RF<ta ble_number><cal_fac tor>%
ta ble
number: 0 to 9
cal fac tor: 50 to 150
%: terminator
Re marks: Set the ref er ence cal fac tor value for a ta ble. Since the ML2400A stores
the cal fac tor ta ble in for ma tion in the sen sor, this data must be saved to
the sen sor by us ing the ‘EX’ com mand or an ad di tional com mand ‘SV’,
or the data could be lost. The sav ing of the cal fac tor ta ble data to the
sen sor can be done at the end of all up dates to a par ticu lar ta ble.
RH Range hold
Syn tax: RH
Re marks: Hold the power me ter in the cur rent range. The dif fer ences in sen sor
ranges must be taken into ac count when the ML2400 is be ing used to
em u late the HP 437.
ML2400A OM 6-115
GPIB OPERATION HP 437B EMULATION COMMANDS