EasyManua.ls Logo

Anritsu ML2400A Series - Page 237

Anritsu ML2400A Series
275 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Er ror Message Meaning
SENSOR A not fit ted
READ ING SEN SOR A data
Sensor A is required for the cur rent
meas ure ment, but is not fitted.
as per message
SENSOR B not fit ted
READ ING SEN SOR B data
Sen sor B is required for the cur rent
meas ure ment, but is not fitted.
as per message
Printer buffer full
Printer - Check pa per
Printer Error
Try selecting print, when current job finished.
Printer reporting “0ut of paper.
Printer com mu ni ca tion error.
Chan 1 - NO trig ger
Chan 2 - NO trig ger
Wait ing for trig ger on speci fied chan nel for
more than 6 sec onds
In crease PE RIOD to > 6 ms WARN ING: In pro file mode, when us ing
man ual or con tinu ous trig ger, the data col -
lec tion pe riod must be > 6 ms.
Graph Dis play HELD warn ing mes sage
Up dat ing SEN SOR data Pro gram ming SEN SOR EE PROM
Graph Chan nel Off In any graph mode, the se lected dis play
chan nel is off.
Ext Volts ZERO in prog ress as per mes sage
Put sen sor number limit At tempt to read/write sen sor B setup on
ML2407A
Sen sor A cal fac tor er ror Cal fac tor out of range - sen sor A
Sen sor B cal fac tor er ror Cal fac tor out of range - sen sor B
PUT cur rent mode limit Pre sent con figu ra tion will not al low re -
quested change.
Crest/Peak NOT available CDMA av er age value is less than ap prox i -
mately –27 dBm. Crest and Peak read ings
are not avail able be low this level.
In ter nal er ror (P6) En tered nu meric value out of range
In ter nal er ror (P10) Re quested ac tion NOT al lowed
In ter nal er ror (P20) Cal fac tor edit - du pli cate fre quency en tered
In ter nal er ror (P21) Can not de lete last cal fac tor data pair in se -
lected ta ble, or can not add cal fac tor data
pair in se lected table
In ter nal er ror (P22) Can not de lete 50 MHz cal fac tor in se lected
ta ble
In ter nal er ror (P23) Non valid cal fac tor ta ble
SYSTEM ERROR MESSAGES APPENDIX A