Minute Enter the minute (0-59) using the Up/Down arrow key or
the keypad. Press ENTER when data entry is complete or ESCAPE
to restore the previous value.
q
Month Enter the month (1-12) using the Up/Down arrow key or the
keypad. Press ENTER when data entry is complete or ESCAPE to
restore the previous value.
q
Day Enter the day using the Up/Down arrow key or the keypad.
Press ENTER when data entry is complete or ESCAPE to restore the
previous value.
q
Year Enter the year (2003-2036) using the Up/Down arrow key or
the keypad. Press ENTER when data entry is complete or ESCAPE
to restore the previous value.
q
Back Returns to the top-level SYS menu.
q
Printer Displays a menu of supported printers. Use the Up/Down arrow
key and ENTER key to make the selection.
q
Units Select the unit of measurement (metric or English).
q
Change Date Format Toggles the date format between
MM/DD/YYYY, DD/MM/YYYY, and YYYY/MM/DD.
q
Back Returns to the top-level SYS menu.
q
Application Options Application options displays options applicable to par
-
ticular mode.
q
External Ref Freq Set the external reference frequency from 2 MHz to
20 MHz.
q
Bias Tee Selects Bias Tee on or off.
q
Impedance Sets the input impedance of the instrument
q
50W Sets the input impedance of the instrument to 50W, which is
the default setting.
q
75W Anritsu 12N50-75B Displays measurements as if the input
impedance was 75W. The actual input impedance is still 50W, but the
displayed information is transformed to the equivalent 75W value
based on the offset value specified by the adapter used in the measure
-
ment. This setting is commonly used for cable TV-related measure
-
ments. Pressing this soft key will also compensate 7.5 dB for the loss
of the Anritsu 12N50-75B adapter.
q
Other adapter Offset This key allows the user to enter other adapter
offset values. Enter the desired loss using the keypad or Up/Down ar
-
row key and press the ENTER key to accept.
q
Back Returns to the previous menu.
q
Back Returns to the top-level SYS menu.
q
Self Test Start an instrument self test.
q
Status In cable and antenna analyzer freq or dist measurement mode, dis
-
plays the current instrument status, including calibration status, temperature,
and battery charge state. Press ESCAPE to return to operation.
q
GPS Opens the GPS soft key menu.
q
Press the GPS On/Off soft key to turn the GPS feature on or off.
q
Press the Location soft key to view the latitude, longitude and altitude in
-
formation and UTC timing.
Chapter 2 Functions and Operations
2-59