Chapter 1 Before Starting the Installation
Planning the Installation
6 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
~
02:00:00
7300/7500
Perform the pure dye calibration (Chapter 6):
a. Prepare the
pure dye plates.
b. Create a plate document.
c. Perform the
pure dye calibration.
d. Analyze the pure dye calibration data.
e. Evaluate pure dye calibration spectra.
~
01:20:00
~
01:00:00
Verify instrument performance (Chapter 7):
a. Prepare the TaqMan
®
RNase P verification plate.
b. Create a plate document.
c. Perform the
verification run.
d. Analyze the TaqMan
®
RNase P plate data.
7500 Fast
~
00:40:00
7500/7500 Fast7300
Note: You must be present at the instrument for approximately 5 minutes of the total run time.
7500 Fast
Recommended Workflow (continued)
See Note below
Perform the optical calibration (7500/7500 Fast systems only,
Chapter 5):
a. Create a plate document.
b. Perform the optical calibration.
c. Analyze the optical calibration data.
~
00:10:00
Perform the background calibration (Chapter 5):
a. Prepare the
background calibration plate.
b. Create a plate document.
c. Perform the background calibration.
d. Analyze the background data.
~
00:30:00
Time Required Task

Table of Contents

Other manuals for Applied Biosystems 7500 Fast

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 Fast and is the answer not in the manual?

Applied Biosystems 7500 Fast Specifications

General IconGeneral
BrandApplied Biosystems
Model7500 Fast
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals