Chapter 5 Performing the Background Calibration and Optical Calibration
Troubleshooting
64 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Troubleshooting
Troubleshooting – Background Calibration
Condition: Cannot extract the data (background calibration failed)
Signals that exceed 72,000 fluorescent standard units (FSU) are considered beyond the limit of normal background
fluorescence for a 7300/7500/7500 Fast instrument. Such signals may indicate that either the background or the sample
block contains fluorescent contaminants. Common contaminants include: ink residue from permanent pens, powder from
disposable gloves, and dust.
To determine the source and location of the contamination:
1. In the plate document for the calibration:
a. Select the Results tab.
b. Select the Spectra tab.
2. Select all wells of the plate document.
3. Inspect the raw background data for an irregular spectral peak
or peaks.
Wells producing raw spectra that exceed 72,000 FSU are
considered irregular and could be contaminated.
a
b
Click to select
all wells
Possible
contamination

Table of Contents

Other manuals for Applied Biosystems 7500 Fast

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 Fast and is the answer not in the manual?

Applied Biosystems 7500 Fast Specifications

General IconGeneral
BrandApplied Biosystems
Model7500 Fast
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals