42 www.arkrayusa.com
Repeat the same procedure for the
lancing device (step
1
to step
6
).
* After the pre-cleaning and disinfection procedure, the
control solution should be tested to confirm that the meter
worksproperlybeforeusingthemeter.Controlsolution
testsshouldbeperformedwithtwodifferentlevelsof
solutions(ControlLevel1and2).Verifythatthetestresults
arewithintherangeprintedontheteststripvial.Seepages
19-20forhowtodoacontrolsolutiontest.
7
Note: If any of the following deterioration signs appear after
pre-cleaning or disinfecting, please stop using the system
and contact Customer Service, 800.566.8558, 24 hours a
day, 7 days a week, US and Canada.
• When the inscriptions on the exterior of the meter (or
lancing device) have been removed.
• When the color of the meter (or lancing device) has
changed or faded.
• When cracks or roughness develop on the meter (or
lancing device).
• When a part of the segment on the meter display does not
flash.
• When control solution test results do not fall within the
stated range on the test strip vial.