Installation
29
28. With the lid of the pump still open, set the speed of the pump to 30% by
typing this value into the display box on the software screen. The pump
should begin to operate.
29. With the pump running, close the pump lid. Water should now be drawn along
the tube from the large process vessel, through the pump and then through
the differential pressure sensor set within the manifold block (in front of the
large process vessel).
30. Check that there is a positive reading from the differential pressure sensor
P3. Increase the speed of Pump B, and check that the differential pressure
reading also increases. Reduce the speed of Pump A to 0.