92
10.3.5. Calibration of DC 10mA Range
Press the key, the program will advance to the DC 10mA range calibration of the hipot test.
The display will show:
DC 10mA, 100K
T E S T t o C a l i b r a t e
Connect a resistor about 100K/10W in series with a DC standard Ammeter which can
measure up to 10mA to the output leads. The Ammeter should be connected to the return lead,
then press the TEST button, the program will automatically generate an output of about
1000V/10mA and the display will show:
Current = mA
Enter Standard I-out
Use the Numeric keys to enter the standard value of current, unit mA, and then press the
ENTER key to store the standard value of DC 10mA range for calibration and the display will
show:
DC 10mA, 100K OK
T E S T t o C a l i b r a t e
10.3.6. Calibration of AC 3.5mA Range
Press the key, the program will advance to the AC 3.5mA range calibration of the hipot test.
The display will show:
AC 3.5mA, 100K
T E S T t o C a l i b r a t e
Connect a resistor about 100K/10W in series with a AC standard Ammeter which can
measure up to 3mA to the output leads. The Ammeter should be connected to the return lead,
then press the TEST button, the program will automatically generate an output on the output
connectors about 300V/3mA and the display will show:
Current = mA
Enter Standard I-out
Use the Numeric keys to enter the standard value of current, unit mA, and then press the
ENTER key to store the standard value of AC 3.5mA range of the hipot test and the display
will show: