59
GND Dwell XXX.X s
MXX-X - - . - - A - - - m
During the Ground Bond test, when the values are being updated in real time, the display
will show:
GND Dwell XXX.X s
MXX-X XX.XX A XXX m
If the DUT failed the HI-Limit of the Ground Bond test and the ground continuity
resistance is within the metering range, the display will show:
GND HI-Limit XXX.X s
MXX-X XX.XX A XXX m
If the DUT failed the HI-Limit of the Ground Bond test and the ground continuity
resistance is beyond the metering range, the display will show:
GND HI-Limit XXX.X s
MXX-X 0.00 A 600 m
If the DUT resistance falls below the LO-Limit of the Ground Bond test, the display will
show:
GND LO-Limit XXX.X s
MXX-X XX.XX A XXX m
When the DUT passed the Ground Bond test when the test process is complete, the
display will show:
GND Pass XXX.X s
MXX-X XX.XX A XXX m
5.7.2. AC Withstand Displayed Messages
If the test in process is Aborted with the RESET button or remote control, the display will
show:
ACW Abort XXX.X s
MXX-X X.XX KV XX.XX mA
If the test in process is Aborted with the RESET button or remote control before the
meter readings are taken, the display will show:
ACW Abort XXX.X s
or
ACW Abort XXX.X s