60
MXX-X - . - - KV - - . - - mA
MXX-X X.XX KV - - . - - mA
At the beginning of AC Withstand Voltage test when the voltage begins to ramp but
before the meter readings are taken, the display will show:
ACW Ramp XXX.X s
MXX-X - . - - KV - - . - - mA
During the AC Withstand Voltage test when the values are being updated in real time
during the ramp cycle, the display will show:
ACW Ramp XXX.X s
MXX-X X.XX KV XX.XX mA
During the AC Withstand Voltage test when the values are being updated in real time
during the dwell cycle, the display will show:
ACW Dwell XXX.X s
MXX-X X.XX KV XX.XX mA
If the ramp time is very short and the program has not read the meter readings, the
display will show:
ACW Dwell XXX.X s
MXX-X - . - - KV - - . - - mA
If the DUT current exceeds the HI-Limit of AC Withstand Voltage test and the leakage
current is within the metering range, the display will show:
ACW HI-Limit XXX.X s
MXX-X X.XX KV XX.XX mA
If the DUT current exceeds the HI-Limit of AC Withstand Voltage test and the leakage
current is not within the metering range, the display will show:
ACW HI-Limit XXX.X s
MXX-X X.XX KV 40 mA
If the DUT current is well beyond the metering range of AC Withstand Voltage test the
instrument assumes that the failure is due to a short circuit, the display will show:
ACW Short XXX.X s
MXX-X - . - - KV 40 mA