GROUND BOND TEST MODE Model 7704 only
Output Voltage
Range:
3.00 - 8.00 Volts AC
(Open Circuit Limit)
Resolution:
0.01 volt/step
Accuracy:
(2 % of Setting + 0.03V ) O.C. Condition
Output Frequency
Range:
60 or 50 Hz, User Selection
Accuracy:
1%
Output Current
Range:
3.00 - 30.00 Amps AC
Resolution:
0.01 Amp/step
Accuracy :
(2 % of Setting + 0.02 A)
Current Display
Range:
0.00 - 30.00 Amps
Resolution:
0.01 Amp/step
Accuracy:
(3 % of Reading + 0.03 A)
Resistance Display
Range:
0 - 600 m
Resolution:
1 m/step
Accuracy:
(2 % of Reading + 2 m )
HI-Limit
Range:
0 - 600 m for 3 - 10 A
0 - 150 m for 3 - 30 A
Resolution:
1 m/step
Accuracy:
(2 % of Setting + 2 m )
LO-Limit
Range:
0 - 600 m for 3 - 10 A
0 - 150 m for 3 - 30 A
Resolution:
1 m/step
Accuracy:
(2 % of Setting + 2 m )
Dwell Timer
Range:
0, 0.5 - 999.9 sec (0 = Constant)
Resolution:
0.1 sec/step
Accuracy:
(0.1% + 0.05 sec)
Milliohm Offset
Max. Offset
Capability:
200 m
Resolution:
1 m / step
Accuracy:
(2 % of Setting + 2 m )