209
lower right screen will change to the text indicated in the table above (Cal. Active).
Read the measurement from your standard and enter it using the numeric keypad.
You may now store the new calibration number by pressing the ENTER key or escape
by pressing the EXIT key or the RESET button.
Calibration of DC Continuity
The load/meter screen and prompt screens will contain the following text for this
calibration point:
Upper Right Screen
Load/Meter
Lower Right Screen
Prompt
Lower Right Screen
Prompt (Cal. Active)
Step 1: Short the Current
and Return leads
together.
When the leads are
shorted, press TEST to
start the calibration
process.
Automatic Calibration
Processing.
Step 2: Connect the
standard 10 load from
Current to Return.
When the load is
connected, press TEST to
start the calibration
process.
Automatic Calibration
Processing.
Step 3: Connect the
standard 100 load from
Current to Return.
When the load is
connected, press TEST to
start the calibration
process.
Automatic Calibration
Processing.
Step 4: Connect the
standard 1000 load from
Current to Return.
When the load is
connected, press TEST to
start the calibration
process.
Automatic Calibration
Processing.
Step 5: Connect the
standard 10000 load
from Current to Return.
When the load is
connected, press TEST to
start the calibration
process.
Automatic Calibration
Processing.
Press the TEST button to activate the calibration. After the calibration is activated, the
lower right screen will change to the text indicated in the table above (Cal. Active).
This calibration is automatic and does not require data entry.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research OMNIA II 8204 and is the answer not in the manual?

Associated Research OMNIA II 8204 Specifications

General IconGeneral
BrandAssociated Research
ModelOMNIA II 8204
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals