INSULATION RESISTANCE TEST MODE
Range: 30 – 1000 VDC
Resolution: 1 V
Accuracy: (2% of setting + 2 counts)
Range: 0 – 1000 V
Resolution: 1 V
Accuracy: (2% of reading + 2 counts)
0.05M – 50000 M (4 Digit, Auto Ranging)
50 – 499V
0.05M – 999.9M, (7% of reading +2 counts)
500 – 1000V
0.05M – 999.9M, (2% of reading +2 counts)
1000M – 9999M (5% of reading +2 counts)
10000M – 50000M, (15% of reading +2
counts)
Range: 0.000 – 3.500A or Auto Set
Range: 0.05M – 99.99M
Resolution: 0.01M
Range: 100.0M – 999.9M
Resolution: 0.1M
Range: 1000M – 50000M
Resolution: 1M
(HI – Limit: 0 = OFF)
Accuracy: Same as Resistance Display Accuracy
Range: Ramp-Up: 0.1 – 999.9 sec
Ramp-Down: 0.0, 1.0–999.9 sec
Resolution: 0.1 sec
Accuracy: (0.1% + 0.05 sec)
Range: 0.5 – 999.9 sec (0 = Continuous)
Resolution: 0.1 sec
Accuracy: (0.1% + 0.05 sec)
GFI Trip Current: 0.4mA - 5.0mA
HV Shut Down Speed: 1mS