18
GROUND BOND TEST MODE
Output Voltage
(Open Circuit Limit)
Range: 3.00 8.00 VAC
Resolution: 0.01 V
Accuracy: (2 % of setting + 0.03V) O.C. Condition
Output Frequency
Range: 60 or 50 Hz, user selectable
Accuracy: 0.1%
Output Current
Range: 1.00 40.00 A
Resolution: 0.01 A
Accuracy: (2 % of setting + 0.02 A)
Output Regulation
Accuracy: (1% of output + 0.02A)
Within maximum load limits, and over input voltage range.
Maximum Loading
1.00 10.00A, 0 600m
10.01 30.00A, 0 200m
30.01 40.00A, 0 150m
Current Display
Range: 0.00 40.00 A
Resolution: 0.01 A
Accuracy: (3 % of setting + 0.03 A)
HI and LO-Limits
Range: 0 150 m for 30.01 40.00 Amps
0 200 m for 10.01 30.00 Amps
0 600 m for 1.00 10.00 Amps
Resolution: 1 m
Accuracy: Same as Ohmmeter Display
Ohmmeter Display
Range: 0 150 m for 30.01 40.00 Amps
0 200 m for 10.01 30.00 Amps
0 600 m for 6.00 10.00 Amps
Resolution: 1 m
Accuracy: (2 of reading + 2 m)
Range: 0 600 m for 1.00 5.99 Amps
Resolution: 1 m
Accuracy: (3 of reading + 3 m)
Dwell Timer
Range: 0.5 999.9 sec (0 = Continuous)
Resolution: 0.1 sec
Accuracy: (0.1% + 0.05 sec)
Milliohm Offset
Range: 0 200m
Resolution: 1 m
Accuracy: (2 % of setting + 2 m)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research OMNIA II 8204 and is the answer not in the manual?

Associated Research OMNIA II 8204 Specifications

General IconGeneral
BrandAssociated Research
ModelOMNIA II 8204
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals