Output Voltage
(Open Circuit Limit)
Range: 3.00 – 8.00 VAC
Resolution: 0.01 V
Accuracy: (2 % of setting + 0.03V) O.C. Condition
Range: 60 or 50 Hz, user selectable
Accuracy: 0.1%
Range: 1.00 – 40.00 A
Resolution: 0.01 A
Accuracy: (2 % of setting + 0.02 A)
Accuracy: (1% of output + 0.02A)
Within maximum load limits, and over input voltage range.
1.00 – 10.00A, 0 – 600m
10.01 – 30.00A, 0 – 200m
30.01 – 40.00A, 0 – 150m
Range: 0.00 – 40.00 A
Resolution: 0.01 A
Accuracy: (3 % of setting + 0.03 A)
Range: 0 – 150 m for 30.01 – 40.00 Amps
0 – 200 m for 10.01 – 30.00 Amps
0 – 600 m for 1.00 – 10.00 Amps
Resolution: 1 m
Accuracy: Same as Ohmmeter Display
Range: 0 – 150 m for 30.01 – 40.00 Amps
0 – 200 m for 10.01 – 30.00 Amps
0 – 600 m for 6.00 – 10.00 Amps
Resolution: 1 m
Accuracy: (2 of reading + 2 m)
Range: 0 – 600 m for 1.00 – 5.99 Amps
Resolution: 1 m
Accuracy: (3 of reading + 3 m)
Range: 0.5 – 999.9 sec (0 = Continuous)
Resolution: 0.1 sec
Accuracy: (0.1% + 0.05 sec)
Range: 0 – 200m
Resolution: 1 m
Accuracy: (2 % of setting + 2 m)