19
CONTINUITY TEST MODE
Output Current
DC 0.1A 0.01A Total Resistance*:0.00 12.0
DC 0.01A 0.001A Total Resistance*:12.1 120
DC 0.001A 0.0001A Total Resistance*:121 1200
DC 0.0001A 0.00001A Total Resistance*:1201 10000
Resistance
Display
Range 1: 0.00 10.00
Resolution: 0.01
Accuracy: (1 % of reading + 3 counts)
Range 2: 10.1 100.0
Resolution: 0.1
Accuracy: (1 % of reading + 3 counts)
Range 3: 101 1000
Resolution: 1
Accuracy: (1 % of reading + 3 counts)
Range 4: 1001 10000
Resolution: 1
Accuracy: (1 % of reading + 10 counts)
HI and LO-Limits
Range 1: 0.00 10.00
Resolution: 0.01
Accuracy: (1 % of reading + 3 counts)
Range 2: 10.1 100.0
Resolution: 0.1
Accuracy: (1 % of reading + 3 counts)
Range 3: 101 1000
Resolution: 1
Accuracy: (1 % of reading + 3 counts)
Range 4: 1001 10000
Resolution: 1
Accuracy: (1 % of reading + 10 counts)
(Max Limit: 0 = OFF)
Dwell Timer
Range: 0.0, 0.3 999.9 sec (0 = Continuous)
Resolution: 0.1 sec
Accuracy: (0.1% + 0.05 sec)
Milliohm Offset
Range: 0.00 10.00
Resolution: 0.01
Accuracy: (1 % of reading + 0.03 )

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research OMNIA II 8204 and is the answer not in the manual?

Associated Research OMNIA II 8204 Specifications

General IconGeneral
BrandAssociated Research
ModelOMNIA II 8204
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals