3. Device installation and start-up
page 15 of 47 issue date 24.02.2021, release 11.1
For program groups, the following options are available:
Preset programs
Preset sequences only for electrotherapy
User programs
User sequences only for electrotherapy
I/t curve only for electrotherapy, channel A
Voll acupuncture programs only for laser point probes
Nogier acupuncture programs only for laser point probes
For medical fields, instead of the preset treatment programs and sequences the following options classified by
medical nomenclature are available:
Orthopedics
Sports medicine
Aesthetic medicine
Rheumatology
Neurology
Urology
Dermatology
Angiology
The classification of preset treatment programs and sequences into the above mentioned categories do not limit
their applications in other fields, according to the knowledge and experience of doctors and physiotherapists.
3.2.4.3 US units
This function allows for selection of indication units for ultrasound therapy and combined therapy:
Power watts [W]
Power density Watt per square centimeter [W/cm
2
]
In order to set the appropriate option, select or clear the selection field by its pressing.
3.2.4.4 US head sensitivity
For advanced users only!
It is possible to modify sensitivity of ultrasound heads in terms of no contact detection. According to treatment
specifics, the sensitivity can be improved or reduced beyond default settings.
To change it you need to:
Step Description
1.
2. Select the ultrasound head type.
3.