3. Device installation and start-up
PHG 700C / 701C User Manual Page 16 of 47
Main features of sensitivity settings are listed below.
Sensitivity settings Advantages Disadvantages
Default Default manufacturer settings
High
forced use of precise treatment technique,
i.e. for training purposes
protection of the ultrasound head against
too fast usage
improved sensitivity on
load’s changes
treatment timer stops more
frequently
Low
reduced sensitivity on load’s changes
facilitated treatment performing on small
parts of the body
facilitated treatment performing on areas
including bones, e.g. hands
facilitated phonophoresis treatment
performing
treatment timer does not stop frequently
rapid usage of ultrasound
head possible overheat
reduced comfort of patient’s
sensations possible increase
of heat feeling in the tissue
3.2.4.5 US head bad contact signaling
This function allows you to set the time after which in case of US head weak contact the treatment will be
interrupted.
Option Explanation
Default (after 5 sec.)
The treatment will be interrupted after five seconds when the US head
will lose the acceptable level of contact quality with the patient’s body.
Delayed (after 10 sec.)
The treatment will be interrupted after ten seconds when the US head
will lose the acceptable level of contact quality with the patient’s body.
In order to set the appropriate option, select or clear the selection field by its pressing.
3.2.4.6 Batterymodel 701C
This function allows the user to set the battery charging mode.
Option Explanation
Charge battery up to 80%
Reducing the battery power level extends its lifetime at the expense of
shorter working hours in case of mains power failure.
Charge battery up to 100%
Maximum battery working hours. Keep in mind that full battery
charging shorten its life.
3.2.5 Service
3.2.5.1 Electrodes test
This function allows the user to quickly check the usage status of electrodes applied in electrotherapy treatments.
It can also be successfully used to assess the continuity of patient’s cables.
In order to exercise the option, follow the instructions shown on the display. Detailed description of
the function is described in the PhysioGo Technical description manual.
3.2.5.2 US head calibration
This function allows the user to change the ultrasound head parameters in case of frequent contact quality
problems caused by the frequency changes of ultrasound transducers characteristics.