34
Cleaning/disinfection
ATTENTION
Damage to the device due to cleaning with xatives!
Stains cannot be removed permanently.
• Do not use aldehydes before and during cleaning.
• Donotexposetheproducttotemperatures>40°C/104°Fbeforeorduringclean-
ing.
ATTENTION
Unsuitable accessories!
Risk of damage to the product.
• Only use lint-free, soft cloths.
• Alwaysusedemineralisedwaterforthenalrinse.
• Follow the corresponding operating instructions of all aids and devices used.
ATTENTION
Unsuitable cleaning agents and disinfectants.
The product may become damaged.
• Do not use any process chemicals containing the following ingredients on plastic
parts:
• Chloramides or phenol derivatives
• Do not use any process chemicals containing the following ingredients on stainless
steel:
• Organic or inorganic bases
• Alkaline solutions
• Do not use any process chemicals which contain the following ingredients on
padding:
• Polish or wax polish
• Chemical cleaning agents
• Oils, greases or alcohol
ATTENTION
Incorrect mechanical cleaning and disinfection.
Corrosion due to moisture.
Removetheproductsimmediatelyoncetheprogrammeisnished.
• Prior to using the device on a new patient, all the parts that come into contact with
thesuctionmaterial(canister,lid,overowprotectionandhose)mustalwaysbe
cleanedandthendisinfected.Disposableitems,suchasthebacterialandvirallter,
catheters,ngertipsetc.,mustbeexchanged.Seetheindividualcleaninginstruc-
tions! (“5.3 Cleaning instructions” on page 38).
• If the suction device is only used for one patient, the device and its accessories
should be cleaned and disinfected for hygiene reasons. See the individual cleaning
instructions! (“5.3 Cleaning instructions” on page 38)