48
Technical data
Max. vacuum at MSL -76kPa(-760mbar;-570mmHg)*,or
80% of daily air pressure
Vacuum adjustment Innitelyvariablevacuumregulator
(mechanical regulation valve)
Vacuum display Vacuum gauge (precision class: 2.5)
-1...0bar(±2.5%ofthenalvalue)
(mmHg, bar, kPA)
Pump Swing piston pump
Available collection canister
Reusable secretion canister:
1 l DDS secretion canister
Interfaces for the use of disposable
systems:
1 l
Sound pressure level 56dB(A)
Operating mode Interval operation
(max. 10 minutes “ON”; min. 30 minutes
“OFF”)
Protective earth conductor resistance -
Earth leakage current -
Touch current -
Patient leakage current -
Ambient conditions:
Transport/storage
• Temperature -30 to +50 °C
• Air humidity without condensation 5 to 90%
• Air pressure 700 to 1060 hPa
Ambient conditions: operation
• Temperature +10 to +35 °C
• Air humidity without condensation 20 to 80 %
• Air pressure 700 to 1060 hPa
Contamination level 2
Overvoltage category II
Dimensions (H x W x D) 250x255x180mm
Weight 4kg
Periodic tests Recommended: testing every 24 months
Electric shock protection class (in acc. with
EN 60601-1)
II
Classicationofapplicationparts
Type BF applied parts
Type of protection IP21
CE mark