M2000 register write properties Page 28/30
MID_M2000_BA_MODBUS_02_1805
150
Full pipe
calibration [volts]
0x0122 0 3 -- YES
151 Empty pipe mode 0x0124 0 5
0 = Off
1 = On
2-5 = Reserved
YES
152
Command action
request
0x0125 0 0
0 = No action
1 = Save RAM to
EEPROM
2 = Load EEPROM to
RAM
3 - 5 = Reserved
6 = Save totalizers to
EEPROM
7 = Clear T1 or T+
8 = Clear T2 or T-
9 = Clear TN
10 = Clear all totalizer
11 = Clear port A
diagnostics
12 = Clear port B
diagnostics
13 = Clear all error
counts
14 - 20 = Reserved
21 = Clear power off
totalizer
22 = Reserved
23 = Remote reset
24 = Restore factory
Calibration data
25 = Reserved
26 = Reserved
27 = Reset batch
amount
28 = Remote logoff
29 = Save RAM to
EEPROM
(port B only)
YES
153
Flow simulation
0x012A
-100
100
0xFF80 = Disable
NO
** Command action code 0x01 should be issued to the command action request register
immediately following any non volatile register uptdate**

Other manuals for Badger Meter ModMAG M2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Badger Meter ModMAG M2000 and is the answer not in the manual?

Badger Meter ModMAG M2000 Specifications

General IconGeneral
BrandBadger Meter
ModelModMAG M2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals