8. Important information
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment
guidance
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5%U
T
1
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60%dipinU
T
)
for 5 cycles
70% U
T
(30%dipinU
T
)
for 25 cycles
<5%U
T
(>95% dip in
U
T
) for 5s
<5%U
T
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5%U
T
(>95% dip in
U
T
)for5s
Mains power quality
should by that of a typical
commercial or hospital
environment. If the user
of the Coronis Fusion
10MP requires continued
operation during power
mains interruptions, it is
recommended that the
Coronis Fusion 10MP
be powered from an
uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic eld
IEC 61000-4-8
3A/m
Not applicable
Power frequency
magnetic elds should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms
150 kHz to 80 MHz
3V/m
80 MHz to 2.5 GHz
3V
3V/m
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the Coronis Fusion
10MP, including cables,
than the recommended
separation distance
calculated from the
equation applicable
to the frequency
of the transmitter.
Recommended
separation distance
d=1.2P
d=1.2P 80 MHz to 800
MHz
d=2.3P 800 MHz to 2.5
Ghz
Where P is the maximum
output power rating
of the transmitter in
watts (W) according
to the transmitter
manufacturer and d
is the recommended
separation distance in
meters (m).
Field strengths from
xed RF transmitters,
as determined by an
electromagnetic site
survey,
2
should be less
1. is the a.c. mains voltage prior to application of the test level.
K5902031 CORONIS FUSION 10MP 28/08/2013 43

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Barco MDCG-10130 and is the answer not in the manual?

Barco MDCG-10130 Specifications

General IconGeneral
BrandBarco
ModelMDCG-10130
CategoryMonitor
LanguageEnglish

Related product manuals