CAMP/MG208/20-0006 AK 98 Independent Home Dialysis System 12
Lining the machine – Heparin Syringe Pump - cont’d
5. Insert the plastic collar (situated at the end of the cylinder
part of the syringe) into the groove of the heparin pump (1).
6. Bring back heparin pump piston in line with the head the
syringe plunger and insert the head of the syringe plunger in
the groove next to the locking wheel (3).
7. Turn the locking wheel upwards until you feel resistance (4).
8. Check that the syringe is properly attached by lightly pulling
the plunger (5).
Note: The heparin pump will start to run at treatment start when
blood is detected in the venous bloodline. The heparin pump is
pre-set and will automatically stop ___min before the end of
treatment.