CAMP/MG208/20-0006 AK 98 Independent Home Dialysis System 25
End of Treatment and Disconnection (Section 4.3.1)
When the remaining treatment time on the Time
Display is 0:00, an attention alarm appears.
The UF rate will be automatically reduced to min UF rate as
the set UF volume is achieved.
The time shown on the Time Display will continue to count
down but will show negative figures.
1. Before the end of treatment, take note of the following parameters:
Single Needle: Press the Blood Path button.
Single Needle: Press the single needle option to deactivate the single
needle function (Section 5.8).