CAMP/MG208/20-0006 AK 98 Independent Home Dialysis System 19
Setting Parameters – cont’d
Single Needle: After priming is completed, check the priming fluid
level in the expansion chamber is around the groove and return it to
the upright position.
Press the Blood button and open the Blood menu. Press the Single
Needle tab and set a minimum stroke volume limit, and then activate
single needle. The machine is now ready for patient connection.