CAMP/MG208/20-0006 AK 98 Independent Home Dialysis System 27
7. When rinse back is complete, the blood pump will stop automatically
when blood is no longer detected in the venous line or when preset
volume has been achieved.
(Extra rinse back can be repeated as necessary by pressing EXTRA
RINSE BACK).
When rinse back is complete,
8. Clamp the venous bloodline and access.
9. Disconnect the venous line from the access as per hospital policy.
End of Treatment and Disconnection – cont’d