6 General Pump Options
6-2 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 071960491
General Pump Options
The configurable option settings listed in Table 6-1 affect the operation
of the pump regardless of the selected PERSONALITY feature set or
label.
Table 6-1 General Pump Options
Option Available Setting Factory Setting
PERSONALITY
Configuration Settings
One permanent and up to eight custom PERSONALITY
feature sets.
Available (Enable)
Unavailable (Disable)
Permanent Settings
PERSONALITY
feature set Enabled
Custom Label List Up to 436 custom labels can be configured. After custom labels
have been configured, they can then be configured with
programming limits using the COLLEAGUE GUARDIAN
feature. Custom labels can then be enabled or disabled
individually for each PERSONALITY feature set.
No custom labels
defined
COLLEAGUE GUARDIAN
Feature
For all predefined medication labels and custom labels in the
pump’s label library, programming limits can be configured
which will result in warnings to the clinician if they try to
program parameters which exceed the predefined limits for that
label configured by the care site.
No Entry for any
labels
Time Setting
(Real time entered in hours
and minutes)
Hours: Minutes Central Standard
Time
Date Setting Month/Day/Year Current date for
U.S. (CST) in
MMDDYY format
Rate-Volume Infusion Mode Always available and cannot be disabled. (Enabled) Always enabled
Power On Default
PERSONALITY Settings
All named PERSONALITY feature sets.
Available (Enable)
Unavailable (Disable)
Permanent settings
(1) Enabled