EasyManua.ls Logo

baxter Colleague CXE - Programming an Infusion; Primary Infusions; Programming a Primary Rate-Volume Infusion

baxter Colleague CXE
230 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
071960491 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 4-21
Operating Instructions Primary Infusions
Programming an Infusion
Before starting an infusion, the administration set must be fully loaded.
If an administration set has not been loaded and START is
pressed, a Tube Not Loaded alarm will occur.
If an administration set is still loading and START is pressed, a
Tube Loading in Progress alarm will occur.
See Table 8-1, “Troubleshooting Alarm Messages,” on page 8-13 for
more information about these alarms.
Primary Infusions
Programming a Primary Rate-Volume Infusion
1. From the Main Display screen (Figure 4-24):
The display then changes to the Rate-Volume programming screen
(Figure 4-25). The Rate field is highlighted.
Figure 4-24 Main Display Screen
For single channel pumps: For triple channel pumps:
press the Primary soft key or
the
Rate or Volume keys.
press the desired
Channel
Select
key.
Note:
Channel Stopped
Alert
If a pump channel is powered on and no keys are
pressed for two minutes, a
Channel Stopped alert
occurs. Either continue programming, start the infu-
sion, or place the channel in standby to clear the alert.
!
WARNING
!
There may be periods of no flow for flow rates less than
or equal to 1mL/hr.

Table of Contents