071960491 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 7-3
Maintenance and Storage Maintenance
Battery Care
The pump can be battery-powered in emergency situations and while
transporting patients. The battery charge icon is displayed at all times in
the upper left part of the screen. When the pump is unplugged and
operating on battery power, the battery charge icon alternates with the
Plug In icon, and the approximate battery operating time is displayed
below the icon.
Charge the battery for at least 12 uninterrupted hours or until the battery
operating time displayed under the battery charge icon in the upper left
part of the screen is 3h 15m (for triple channel pumps) or 4h 00m (for
single channel pumps). A complete charge may take longer than
12 hours.
Check Action
Perform as required but recommended after every use.
Housings Clean housing and front panel as recommended in the cleaning instructions in this section. Check for
cracks and large dents.
Labels Clean as recommended in the cleaning instructions.
Check for scratches, cuts or obliterated words.
Power cord Check the cord and connectors for cracks and damage. Do not use the pump if the connectors appear
damaged.
Feet Verify that the feet are free of cuts or deterioration and that they are securely fastened.
Rear housing Verify that there are no loose or missing parts and that connectors and accessories are undamaged.
Contrast and
Volume controls
Verify that both controls are undamaged and rotate freely.
Pole clamp knob
and mechanism
Verify that knob operates freely throughout range of motion. Ensure that pads are present; check that
the pump remains attached to IV pole.
Main Batteries Pump should be plugged in whenever possible to maintain batteries at full charge. If batteries are not
at full charge, recharge them by plugging the pump into a 100-120 VAC, 50/60 Hz or 220-240 VAC,
50/60 Hz power outlet for at least 12 hours. Check that the plug icon is illuminated during this time.
See “Storage,” 7-6 for information on charging batteries for pumps in storage.
Perform as required but recommended every 12 months.
Entire pump Schedule operational checkout by Baxter-trained, qualified personnel or authorized service
representative.
CAUTION
When charging the batteries, ensure the room temperature is between 15° C
(59° F) to 30° C (86° F) to minimize charge time and maximize battery life.