Loading...
*
Squeeze
the
release
lever
and
advance
the
pusher
block
until
the
latch
grasps
the
syringe
plunger
flange
and
the
pusher
block
presses
firmly
against
the
syringe
plunger
as
indicated
in
Figure
2-5.
Г
Baxter
PCA
6
Figure
2-5
»
Latching
the
pusher
block
to
the
syringe
plunger
flange
WARNING
|
Make
sure
the
plunger
flange
is
engaged
by
the
anti-siphon
latch,
otherwise
uncontrolled
emptying
of
the
syringe
may
occur.
Uncontrolled
emptying
of
the
syringe
may
result
in
patient
injury.