10 HTS Customer Care Kits User’s Guide
3 Detach the sample coupler from the cytometer SIT.
a Unscrew the top nut.
b Pull the sample coupler down and away from the SIT, leaving the nut
attached to the sample coupler.
4 Unscrew the captive screw located at the bottom of the HTS support
bracket.
5 Pull the HTS out (about 2 inches) from the support bracket.
6 Remove the probe and the overflow reservoir.
7 Disconnect interlock and interface cables.
8 Disconnect the sheath and waste connectors from the HTS.
9 Disconnect the sampler coupler from the cytometer SIT.
10 Disconnect the door sensor cable.
a Lift the front feet of the HTS just over the front edge of the enclosure
and tilt the unit at a 45° angle toward you for access to the door sensor
cable on the right side.
b Unplug the door sensor cable by pulling back on the ferrule.
c Place the cable over the right side of the enclosure/carrier.