32 HTS Customer Care Kits User’s Guide
Required Kit Components
Pump Syringe Replacement Procedures
If you are replacing the primary pump, use both procedures in this chapter.
If you are replacing the secondary pump, remove the HTS (see Chapter 2),
then use the Replacing the Syringe procedures.
Removing Liquid from the Syringe (For Primary Pump Syringe)
1 Verify that the cytometer and BD FACSDiva software are running.
2 From the Acquisition Dashboard select RUN PLATE.
3 Remove the liquid from the syringe:
a Disconnect the sheath connector from the back of the HTS.
b Disconnect one end of the sheath filter.
c Select HTS > Prime.
d Repeat until the syringe barrel is empty.
e Exit BD FACSDiva software.
4 Turn off the cytometer.
Item number Part number Description Qty
6 344912 Syringe, 500 µL, with black plunger seal 1
7 347306 Lubricant, O-ring, Super O-Lube, ½-
ounce tube
1