32 HTS Customer Care Kits User’s Guide
Required Kit Components
Pump Syringe Replacement Procedures
• If you are replacing the primary pump, use both procedures in this chapter.
• If you are replacing the secondary pump, remove the HTS (see Chapter 2),
then use the Replacing the Syringe procedures.
Removing Liquid from the Syringe (For Primary Pump Syringe)
1 Verify that the cytometer and BD FACSDiva software are running.
2 From the Acquisition Dashboard select RUN PLATE.
3 Remove the liquid from the syringe:
a Disconnect the sheath connector from the back of the HTS.
b Disconnect one end of the sheath filter.
c Select HTS > Prime.
d Repeat until the syringe barrel is empty.
e Exit BD FACSDiva software.
4 Turn off the cytometer.
Item number Part number Description Qty
6 344912 Syringe, 500 µL, with black plunger seal 1
7 347306 Lubricant, O-ring, Super O-Lube, ½-
ounce tube
1