Chapter 3: Wash Station Assembly Replacement 17
3
Wash Station Assembly
Replacement
Perform the wash station assembly maintenance tasks every six months or when
the following symptoms occur:
• The HTS exhibits higher than normal carryover between samples
• The wash station overflows
Required Kit Components
The sample injection tubing (3) connects the injection port/wash station (1) to the
secondary pump valve (not shown).
Item
number
Part
Number
Description Qty
1 336947 Injection port/wash station 1
2 335526 Tubing, injection port 1
3 335451 Tubing assembly, sample injection tubing to
HTS pump
1