Chapter 5: Sample Coupler and Tubing Replacement 29
3 Install the new sample coupler tubing provided in the HTS Customer Care
Kit.
4 Connect the new tubing to the left port in the secondary pump valve, and
tighten the tubing until the fitting is finger tight.
NOTE Valves contain sealing washers in each port. If you over-tighten the
fitting, the sealing washers can be compressed, resulting in a blocked port.
Tighten the fitting until it contacts the sealing washer, and then turn the
fitting an additional 1/6 to 1/4 turn. If leaking is observed, tighten the
fitting no more than an additional 1/8 turn.
5 Install the HTS on the cytometer (see Chapter 2).
Secondary pump valve