34 HTS Customer Care Kits User’s Guide
c Tighten the plunger lock screw, making sure it is securely tightened.
6 Apply a generous amount of O-ring lubricant to the plunger lock screw to
prevent seizing in the future. You can apply the lubricant in the lock lever
hole before placing the plunger in it. This will help keep the saline out of
the whole assembly Do not lubricate the syringe barrel.
7 For the secondary syringe, reinstall the HTS on the cytometer (see
Chapter 2).
Lubricant
Plunger
Plunger
holder