46 HTS Customer Care Kits User’s Guide
4 Submerge the other end of the purge line into the DI water.
5 Using BD FACSDiva software, perform a system prime. Repeat 3 times.
For BD FACS Calibur, use BD Plate Manager software to run the prime.
6 Remove the purge tubing.
7 Reinstall the sheath line.
BD HTS sheath port
HTS purge tubing