Alaris™ GP (Guardrails™) Volumetric Pump
Basic Features
BDDF00535 Issue 1
25/56
Primary Setup
If a secondary infusion has already been setup (See 'Secondary (Piggyback) Infusions' section), then access to the primary infusion setup
is as follows:
1. Press
h to put the Pump ON HOLD.
2. Press the
d button to access the options menu.
3. Select PRIMARY Setup and press the OK softkey to confirm. Make changes to the primary setup as necessary.
Secondary Setup
To setup a secondary infusion refer to 'Secondary (Piggyback) Infusions' section
Setting VTBI over Time
This option allows a specific VTBI and delivery time (Maximum of 24 hours) to be set. The rate necessary to deliver the required volume
within the specified time is calculated and displayed.
1. Stop the infusion. Press
dbutton to access the options menu.
2. Select the SET VTBI OVER TIME option using the
fkeys and press the OK softkey.
3. Adjust the volume to be infused using the
fkeys (Or select BAGS softkey to set the VTBI). When the desired volume has
been reached press the OK softkey.
4. Enter the time over which the volume is to be infused using the
fkeys . The infusion rate will automatically be calculated.
5. Press OK softkey to enter the value or BACK to return to the VTBI.
Adjust Alarm Volume
This option allows adjustment of the volume if enabled.
1. Press the
d button to access the options menu.
2. Select ADJUST ALARM VOLUME using the fkeys and press OK softkey to confirm.
3. Select HIGH, MEDIUM or LOW using the fkeys.
4. Press OK softkey to confirm or QUIT to exit screen.
Event Log
This option allows the event log to be reviewed. The event log holds up to 99,960 events. When the log is full, the oldest events will be
overwritten by the newest occurring events.
1. Press the
d button to access the options menu.
2. Select EVENT LOG using the fkeys and press OK softkey to confirm.
3. View the events using the
f keys.
4. Select BACK to return to the previous screen, if required.
Pump Details
To review Pump information:
1. Press the
d button to access the options menu.
2. Select PUMP DETAILS using the fkeys and press OK softkey to confirm.
3. Review the information and then press the QUIT softkey.
Profile Filter (Only available on the Alaris GP Guardrails Volumetric Pump)
This option allows selected profiles to be filtered from view. It can be enabled/disabled via the Alaris Editor.
1. Press the
d button to access the options menu.
2. Select PROFILE FILTER using the
f keys, press the OK softkey to confirm.
3. Select profile that requires filtering using the
f keys.
4. Press MODIFY to disable profile and press OK to confirm or QUIT to return to main display.
Note: If only one profile is available and all others are disabled in this option, the CONFIRM PROFILE screen will not display upon
start up.
Standby
This option allows the Pump to be placed on standby mode. It can be enabled/disabled via the Alaris Editor.
1. Press the
d button to access the options menu.
2. Select STANDBY using the
f keys, press the OK softkey to confirm.
3. Select CANCEL to return to main display.